Please use this identifier to cite or link to this item: http://dx.doi.org/10.25673/32762
Title: Data Processing and Analysis of Glucose Concentration According to the Immittance Meter
Author(s): Shipilov, Sergey
Klokov, Andrey
Yurchenko, Ekaterina
Zavyalova, Kseniya
Yurchenko, Alexey
Issue Date: 2020
Language: English
Abstract: This article proposes a method for measuring glucose concentration as a possible promising way to create a non-invasive glucometer. The dependence of the complex resistance of a solution of sodium chloride with different concentrations of glucose is investigated. In fist the glucose concentrations were calculated in saline to carry out these measurements. Further we prepared solutions (test samples) with a content of 3, 5, and 10 mmol per liter. We also created a board layout for measurements, that allowed us to vary solutions with different glucose concentrations. Using an immittance meter, we measured the complex resistance as a function of frequency in the range from 25 Hz to 1 MHz. At a medium frequency, we observed a resonance. We also found that at a frequency of 2 kHz there is a clear dependence of the complex resistance on the increase in glucose in solution. The obtained measurement data is of great interest for non-contact monitoring of glucose concentration in biological fluids.
URI: https://opendata.uni-halle.de//handle/1981185920/32946
http://dx.doi.org/10.25673/32762
Open Access: Open access publication
License: In CopyrightIn Copyright
Appears in Collections:International Conference on Applied Innovations in IT (ICAIIT)

Files in This Item:
File Description SizeFormat 
3_4_Shipilov.pdf716.05 kBAdobe PDFThumbnail
View/Open